ILLUMINATION/VIEWING SYSTEM |
Reflectance: di:8°, de:8° (diffuse illumination/8° viewing angle); SCI (specular component included)/SCE (specular component excluded) switchable & simultaneous. Conforms to CIE No.15(2004), ISO 7724/1, ASTM E 1164, DIN 5033 Teil 7, and JIS Z 8722 condition c standards. Transmittance: di:0°, de:0° (diffuse illumination/0° viewing angle). Conforms to CIE No.15(2004), ASTM E 1164, DIN 5033 Teil 7, and JIS Z 8722 condition g standards. |
INTEGRATING SPHERE SIZE |
Ø152 mm (6 inches) |
DETECTOR |
38-element silicon photodiode array |
SPECTRAL SEPARATION DEVICE |
Diffraction grating |
WAVELENGTH RANGE |
360 to 740 nm |
WAVELENGTH PITCH |
10 nm |
HALF BANDWIDTH |
Approximately 14 nm average |
MEASURING RANGE |
0 to 200%; Resolution: 0.001% |
LIGHT SOURCE |
Pulsed xenon arc lamp |
MEASUREMENT/ILLUMINATION AREAS |
Reflectance: Changeable between SAV, MAV, LMAV, and LAV. SAV: 3 × 5 mm measurement / 5 × 7 mm illumination; MAV: Ø8 mm measurement / Ø11 mm illumination; LMAV: Ø16 mm measurement / Ø20 mm illumination; LAV: Ø25.4 mm measurement / Ø28 mm illumination. Transmittance:Approximately Ø20 mm / Ø25 mm. |
REPEATABILITY |
White: Colorimetric values: Standard deviation within ΔEab 0.005; Spectral reflectance: Standard deviation within 0.05% (measured 30 times at 10-second intervals after white calibration). Black: Colorimetric values: Standard deviation within ΔEab 0.02; Spectral reflectance: Standard deviation within 0.02% (measured 30 times at 10-second intervals after white calibration). |
INTER-INSTRUMENT AGREEMENT |
Within ΔE*ab 0.08 (based on average for 12 BCRA Series II color tiles; LAV/SCI. Compared to values measured with a master body under Konica Minolta standard measurement conditions). |
UV ADJUSTMENT |
Computer-controlled; continuously variable, 0.0% to 100.0% (1000 steps). UV setting: UV cutoff filter at 400 nm. |
SAMPLE TEMPERATURE MEASUREMENT |
Accuracy within operating temperature/humidity range: SAV: ±1.2°C; LMAV, MAV: ±0.8°C; LAV: ±0.5°C. |
MEASUREMENT TIME |
SCI or SCE: Approximately 2 seconds; SCI or SCE with sample temperature measurement: Approximately 4.5 seconds; SCI+SCE:Approximately 5 seconds; SCI+SCE with sample temperature measurement: Approximately 5 seconds; Transmittance: Approximately 2 seconds. |
MINIMUM INTERVAL BETWEEN MEASUREMENTS |
SCI or SCE: Approximately 3 seconds; SCI or SCE with sample temperature measurement: Approximately 5 seconds; SCI+SCE:Approximately 6 seconds; SCI+SCE with sample temperature measurement: Approximately 6 seconds; Transmittance: Approximately 3 seconds. |
TRANSMITTANCE CHAMBER |
Maximum sample thickness: Approximately 50 mm; Maximum sample length: Unlimited (no sides when transmittance chamber cover is open). Sample holder (optional) for holding sheet samples or containers of liquid samples can be installed/removed. |
CAMERA VIEWFINDER FUNCTION |
Uses internal camera. Image viewable/copiable using optional software such as SpectraMagic NX2. |
INTERNAL PERFORMANCE CHECK |
Wavelength Analysis & Adjustment (WAA) Technology. |
AMBIENT TEMPERATURE SENSOR |
Yes |
INTERFACE |
USB 2.0 |
POWER |
Dedicated AC adapter |
DIMENSIONS (H × W × D) |
Approximately 307 × 271 × 600 mm |
WEIGHT |
Approximately 20.0 kg |
OPERATING TEMPERATURE/HUMIDITY RANGE |
Temperature: 13 to 33°C; Relative humidity: 80% or less (at 33°C) with no condensation |
STORAGE TEMPERATURE/HUMIDITY RANGE |
Temperature: 0 to 40°C; Relative humidity: 80% or less (at 35°C) with no condensation |
STANDARD ACCESSORIES |
White Calibration Plate; Target Masks (SAV, MAV, LMAV, LAV); Zero Calibration Box; USB Cable (3 m); AC Adapter |
OPTIONAL ACCESSORIES |
Color Data Software SpectraMagic™ NX2; Transmittance Specimen Holder; Cells (Glass: 2 mm, 10 mm, 20 mm); Plastic Cells (2 mm, 10 mm, 20 mm); Transmittance Zero Calibration Plate; Color Plates; Green tile; Dust Cover. |